Nano Dot OSLDs are radiation detectors whose signal output becomes less predictable at high radiation doses. Above 3 Gy, the relationship between dose and signal is no longer proportional, and above 15 Gy, the detectors become less sensitive, so they must be calibrated in separate ranges to measure high doses accurately.

Evidence from Studies

Supporting (1)

0

Community contributions welcome

This study found that the tiny radiation sensors (OSLDs) don’t respond in a straight line when exposed to very high doses of radiation — they get less sensitive above 3 Gy and need special math formulas to measure accurately, just like the claim says.